{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Zurich",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80917",
      "recalling_firm": "Thoratec Switzerland GMBH",
      "address_1": "Technoparkstrasse 1",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: OR, AZ, CA, ID, WA, WI, NY, MA, CT, NJ, ME, RI, IL, MI, TN, OH, IN, KY, IA, AL, FL, NC, GA, SC, VA, MD, DC, PA, WV, DE, MN, MT, NE, CO, NM, UT, AR, MO, MC, TX, OK, LA, MS, ND",
      "recall_number": "Z-0103-2019",
      "product_description": "Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System.",
      "product_quantity": "1569",
      "reason_for_recall": "Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.",
      "recall_initiation_date": "20180905",
      "center_classification_date": "20181012",
      "termination_date": "20210428",
      "report_date": "20181024",
      "code_info": "UDI: 07640135140078. All serial numbers until design change is implemented to make the motor cable more resistant to damage."
    }
  ]
}