  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90871",
      "recalling_firm": "ARROW INTERNATIONAL Inc.",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the state of Florida.",
      "recall_number": "Z-0102-2023",
      "product_description": "ARROW Pressure Injectable Arrowg+ard Blue Advance Two-Lumen Midline Kit, 5.5 Fr, 15 cm, .018 Inch dia., REF ASK-41552-JHVA1.  For peripheral access to the venous system for intravenous therapy, blood sampling, infusion, and pressure injection of contrast media.",
      "product_quantity": "18 devices",
      "reason_for_recall": "Mislabeling:  the corrugate label,  lidstock, banner card, and peel sticker accompanying the affected product incorrectly displays a double lumen 5.5 French catheter (and related priming volumes and kit contents for product code ASK-41552-JHVA1), instead of the packaged single-lumen 4.5 French catheter (and related priming volumes and kit contents for product code ASK-41541-JHVA1).",
      "recall_initiation_date": "20220914",
      "center_classification_date": "20221019",
      "termination_date": "20240119",
      "report_date": "20221026",
      "code_info": "UDI (01)10801902144284(17)230731(11)220504(10)13F22C0757",
      "more_code_info": ""
    }
  ]
}