  "results": [
    {
      "country": "United States",
      "city": "Indianola",
      "address_1": "1 Bayer Dr",
      "reason_for_recall": "Bayer has received reports of loose particulate in the filter when the dust cap is removed. Bayer conducted further testing confirming the dust cap may generate particulate which may potentially enter the fluid path.",
      "address_2": "",
      "product_quantity": "1,142,836 (329,944 units in U.S.)",
      "code_info": "Batch Numbers 14888  72498  72789  74023  75221  17061  72501  72790  74161  75277  20046  72529  72990  74162  75278  20056  72541  73365  74568  75434  70091  72639  73372  74569  75435  72354  72677  73600  74570  75482  72362  72713  73827  74571  75603  72365  72714  73944  74758  75625  72422  72788  74022  75220  75733",
      "center_classification_date": "20181011",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, and WI.  Foreign countries:  UAE, Austria, Australia, Bosnia and Herzegovina, Belgium, Bahrain, Canada, China, Chile, Costa Rica, Czechia, Denmark, German, Spain, Estonia, Egypt, Finland, France, Great Britain, Croatia, Hungary, Israel, Italy, Jordan, Kuwait, Kazakhstan, Lebanon, Lithuania, Morocco, Malta, Mexico, Netherlands, Norway, Oman, Panama, Poland, Portugal, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, and Turkey",
      "state": "PA",
      "product_description": "Codan Filter (Codan catalog number BC 693) used with Medrad Patient Administration Sets (PAS) and Intego PET Infusion Systems    Product Usage:  1.2 micron filter for use with Intravenous Administration Sets",
      "report_date": "20181017",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Bayer Medical Care, Inc.",
      "recall_number": "Z-0097-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80945",
      "termination_date": "20190711",
      "more_code_info": "",
      "recall_initiation_date": "20180907",
      "postal_code": "15051-9702",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}