  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Union City",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86482",
      "recalling_firm": "Mizuho OSI",
      "address_1": "30031 Ahern Ave",
      "address_2": "N/A",
      "postal_code": "94587-1234",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in   states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, SC, TN, TX, VA, WI, and WV.    The countries of Australia, Belgium, Canada, Columbia, Germany, Israel, South Korea, New Zealand, Poland, United Kingdom, and South Africa.",
      "recall_number": "Z-0091-2021",
      "product_description": "Levo Arm Label: LEVO ARM REF 7887-050 MIZUHO OSI    Refurbished Levo Arm Label: LEVO ARM REF 7887-050R  MIZUHO OSI - Product Usage: used with Mizuho OSI spinal surgery tables to provide patient head support and positioning during spinal surgery.",
      "product_quantity": "174 Arms",
      "reason_for_recall": "Due to the potential for fluid ingress into the Arm which could affect the device's electrical system and lead to a failure of the locking mechanism",
      "recall_initiation_date": "20191014",
      "center_classification_date": "20201009",
      "termination_date": "20230927",
      "report_date": "20201021",
      "code_info": "Model Number: 7887-050  UDI/GTIN 00842230104992  Affected Serial Numbers: 201 through 452    Model 7887-050R  UDI/GTIN 00824230108440  Affected Serial Numbers: 234, 235, 248, 255, 341, and 448"
    }
  ]
}