  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Nussloch",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95323",
      "recalling_firm": "LEICA BIOSYSTEMS NUSSLOCH GMBH",
      "address_1": "Heidelberger Str. 17-19",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0088-2025",
      "product_description": "Leica Biosystems Cryostat Model Number CM1900, Material Numbers:  a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples",
      "product_quantity": "25 units",
      "reason_for_recall": "As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.",
      "recall_initiation_date": "20240904",
      "center_classification_date": "20241015",
      "report_date": "20241023",
      "code_info": "a) Material Number 14045227382, UDI/DI 04049188075840, Serial Numbers:  1198121999, 383051996;  b) Material Number 14901900100, UDI/DI 04049188003324, Serial Numbers:  703111997, 1018021999, 1538042001, 2058, 629071997, 2994, 1170111999, 3686102007, 3698;  c) Material Number 14901900700, UDI/DI 04049188003300, Serial Numbers:  1267032000, 2602062004, 1577052001, 1679092001, 2542042004, 1771, 1885032002, 2682092004, 2376112003, 2774122004, 1114, 1749112001, 1997072002;  d) Material Number 1491900UVU2, UDI/DI 04049188002570, Serial Numbers:  3243052006",
      "more_code_info": ""
    }
  ]
}