{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-14",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99807",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Canada.",
      "recall_number": "Z-0086-2027",
      "product_description": "Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS",
      "product_quantity": "1824 units",
      "reason_for_recall": "The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame  tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.",
      "recall_initiation_date": "20260814",
      "center_classification_date": "20261005",
      "report_date": "20261014",
      "code_info": "Lot Numbers: 05425090001 05425120001 05426010001 05426020001 05426030001 05426040001 05426050001"
    }
  ]
}