  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86399",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:  Canada,   Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK,  Algeria, Argentina, Aruba, Australia, Bahrain, Bosnia-Herz., Brazil, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Hong Kong, India, Israel, Japan, Jordan, Kazakhstan, Kuwait, Malaysia, Martinque, Mexico, Monaco, Morocco, Namibia, New Caledonia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Russia, Saudi Arabia, Singapore, S. Africa, S. Korea, Taiwan, Thailand, Ukraine, UAE, Vietnam",
      "recall_number": "Z-0085-2021",
      "product_description": "BenchMark ULTRA Advanced Staining System (IVD)-  Material Number: 05342716001",
      "product_quantity": "2623 (US); 3735 (ex-US)",
      "reason_for_recall": "Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket",
      "recall_initiation_date": "20200831",
      "center_classification_date": "20201008",
      "report_date": "20201014",
      "code_info": "All units"
    }
  ]
}