{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Madison",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95045",
      "recalling_firm": "Datex-Ohmeda, Inc.",
      "address_1": "3030 Ohmeda Dr",
      "address_2": "N/A",
      "postal_code": "53718-6704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Aland Islands, ALBANIA, Algeria, Argentina, Australia, Austria, Azerbaijan, BAHAMAS, Bahrain, Bangladesh, BARBADOS, BELARUS, Belgium, Benin, Bermuda, Bolivia, BONAIRE, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Cameroon, Canada, Chile, CHINA, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, ETHIOPIA, Falkland Islands, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gilbrator, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, ICELAND, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Jamaica, JAPAN, Jordan, Kazakhstan, KENYA, Korea (Republic Of), Korea, Republic of, Kuwait, Laos, LATVIA, Lebanon, Libya, LITHUANIA, Macedonia, Malaysia, Maldives, Malta, MAURITIUS, Mexico, Moldova, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, PAKISTAN, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, PUERTO RICO, Qatar, REUNION, ROMANIA, Russia, San Marino, Saudi Arabia, Senegal, Serbia, SINGAPORE, Slovakia, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, THAILAND, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Uganda, UKRAINE, United Arab Emirates, United Kingdom, Uruguay, UZBEKISTAN, Venezuela, Vietnam, Yemen.",
      "recall_number": "Z-0074-2025",
      "product_description": "Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.",
      "product_quantity": "20,097 units",
      "reason_for_recall": "Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet  (ACGO) port.",
      "recall_initiation_date": "20240712",
      "center_classification_date": "20241018",
      "report_date": "20241030",
      "code_info": "a. Aespire 7100 - GTIN: 00840682102261. Model Number: 1009-9000-000. All Serial Numbers.   b. Aespire 7100 (Variant) - GTIN: 00840682102261. Model Numbers: 1009-9101-000 and 1009-9101-000. All Serial Numbers.    c. Aespire 7100 S/5 Protiva - No GTIN. Model Number: 1009-9004-000. All Serial Numbers.   Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.",
      "more_code_info": ""
    }
  ]
}