  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tucson",
      "state": "AZ",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62516",
      "recalling_firm": "Sunquest Information Systems, Inc.",
      "address_1": "250 S Williams Blvd",
      "address_2": "N/A",
      "postal_code": "85711-4472",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution",
      "recall_number": "Z-0073-2013",
      "product_description": "Sunquest Laboratory :    intended for use by professionals working in a clinical laboratory.",
      "product_quantity": "127 sites",
      "reason_for_recall": "Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.",
      "recall_initiation_date": "20111115",
      "center_classification_date": "20121017",
      "termination_date": "20121019",
      "report_date": "20121024",
      "code_info": "Versions 6.4.0 and later"
    }
  ]
}