  "results": [
    {
      "status": "Terminated",
      "city": "Grand Rapids",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86474",
      "recalling_firm": "Skytron, Div. The KMW Group, Inc",
      "address_1": "5085 Corporate Exchange Blvd SE",
      "address_2": "N/A",
      "postal_code": "49512-5515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "AZ,  MI, MS, NJ, PA,",
      "recall_number": "Z-0071-2021",
      "product_description": "VGDF-SKY (R4-010-06) VGA TO FIBER CONVERTER-Single Format Faceplate, a component of the SkyVision Linx 300 Integration Control System  Model: VGDF-SKY (version 2).  For communication and visual image management in the operating room suite.",
      "product_quantity": "11 units",
      "reason_for_recall": "VGDF (VGA) Faceplate failure results in failure of video signal for SkyVision Linx 300, a secondary display. Video would continue on the primary source device display",
      "recall_initiation_date": "20200911",
      "center_classification_date": "20201008",
      "termination_date": "20240924",
      "report_date": "20201014",
      "code_info": "Serial Numbers:  MXDM200501769  MXDM200501773  MXDM191100421  MXDM191100422  MXDM200200161  MXDM200200162  MXDM200200163  MXDM200200164  MXDM200501770  MXDM200501772  MXDM200501771"
    }
  ]
}