  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Buffalo Grove",
      "address_1": "1700 Leider Ln",
      "reason_for_recall": "These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.",
      "address_2": "",
      "product_quantity": "67 units",
      "code_info": "Serial Numbers: 70-78, 111-112, 120-175",
      "center_classification_date": "20181003",
      "distribution_pattern": "Distributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI.      Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND",
      "state": "IL",
      "product_description": "Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated    Product Usage:  The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.",
      "report_date": "20181010",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Leica Microsystems, Inc.",
      "recall_number": "Z-0064-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80916",
      "termination_date": "20190318",
      "more_code_info": "",
      "recall_initiation_date": "20180613",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}