      "status": "Ongoing",
      "city": "Austell",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99676",
      "recalling_firm": "ASP Global, LLC.",
      "address_1": "7800 Third Flag Pkwy",
      "address_2": "",
      "postal_code": "30168-7657",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.",
      "recall_number": "Z-0061-2027",
      "product_description": "Brand Name: MPR  Product Name:SINGLE USE TOURNIQUET CUFF 8\" 2 LINE (10)  Model/Catalog Number: DT-7108-02Q  Software Version: Not Applicable  Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field.  Component: No",
      "product_quantity": "40",
      "reason_for_recall": "Due to incorrect expiration date on product manufactured between September 2016 to March 2023.  Products are labeled with a 10 year shelf life.  The correct validated shelf life is 5 years.",
      "recall_initiation_date": "20260824",
      "center_classification_date": "20261001",
      "report_date": "20261007",
      "code_info": "Lot Code: Model Number: DT-7108-02Q UDI-DI: UDI UDI was not applicable at time of manufacture.  All Lots produced from 01SEP2016 to 01MAR2023:  S1101STY, R1201STY",
      "more_code_info": ""
    }
  ]
}