  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Redwood City",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95336",
      "recalling_firm": "AURIS HEALTH INC",
      "address_1": "150 Shoreline Dr",
      "address_2": "",
      "postal_code": "94065-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.",
      "recall_number": "Z-0052-2025",
      "product_description": "Monarch Platform (Urology), REF: MON-000007, MON-000007-RFB",
      "product_quantity": "4",
      "reason_for_recall": "Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.",
      "recall_initiation_date": "20240802",
      "center_classification_date": "20241010",
      "report_date": "20241016",
      "code_info": "Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00.  REF/UDI-DI/Serial Numbers: MON-000007/+B634MON0000070/120059, 120133, 210002;  MON-000007-RFB/10810068810889/210001.",
      "more_code_info": ""
    }
  ]
}