  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63157",
      "recalling_firm": "Medtronic Sofamor Danek USA Inc",
      "address_1": "1800 Pyramid Place",
      "address_2": "N/A",
      "postal_code": "38132",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of India, Netherlands, Hong Kong, S. L Korea, Singapore, Canada, Australia, Japan, and Mexico.",
      "recall_number": "Z-0038-2013",
      "product_description": "PREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY:  1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132.    Orthopedic.  This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for  single use, this instrument may be re-used.",
      "product_quantity": "247 units",
      "reason_for_recall": "Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws.  An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.",
      "recall_initiation_date": "20120830",
      "center_classification_date": "20121010",
      "termination_date": "20130815",
      "report_date": "20121017",
      "code_info": "All lots except those starting with SY"
    }
  ]
}