  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 GBC Drive, Mailstop 514",
      "reason_for_recall": "Firm has confirmed highter imprecision with the recalled lots on Beckman Coulter AU(R) Systems.  Also, if the Syva(R) Emit(R) 2000 Phenobartbital Assay (4D019UL) lots D4, D5, D6, and/or D7 as an application on AU(R) Systems is used, the issue may be experienced.  Firm has found that within-run imprecison of the recalled lots when run on an AU(R) System may be higher that what would normally be obs",
      "address_2": "PO BOX 6101",
      "product_quantity": "79,744 cartons",
      "code_info": "Emit(R) 2000 Phenobarbital Assay (OSR4D229), lots 1151, 1167, 1191, and/or 1212 and Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots D4, D5, D6, and D7",
      "center_classification_date": "20121009",
      "distribution_pattern": "United States Nationwide Distribution in the states of: CA, FL, KS, MA, MN, MS, NM, NY, NC, OR, PA, TN, TX, UT, VA, and WA.",
      "state": "DE",
      "product_description": "Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL)    Product Usage: Enzyme n Immunoassay, Phenobarbital",
      "report_date": "20121017",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-0030-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62832",
      "termination_date": "20140827",
      "recall_initiation_date": "20120730",
      "postal_code": "19714-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}