  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Vaxjo",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97535",
      "recalling_firm": "Getinge Disinfection Ab",
      "address_1": "Ljungadalsgatan 11",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.",
      "recall_number": "Z-0016-2026",
      "product_description": "Getinge 88-Series Washer- Disinfector. Model Number: 88-5.",
      "product_quantity": "902 units",
      "reason_for_recall": "Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.",
      "recall_initiation_date": "20250829",
      "center_classification_date": "20251001",
      "report_date": "20251008",
      "code_info": "Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A.  All Lot Numbers with electrical heating."
    }
  ]
}