  "results": [
    {
      "status": "Terminated",
      "city": "Round Lake",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63059",
      "recalling_firm": "Baxter Healthcare Corp.",
      "address_1": "Rt. 120 & Wilson Rd",
      "address_2": "N/A",
      "postal_code": "60073",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and Internationally to Canada, Costa Rica, Australia, New Zealand, Hong Kong, and the United Kingdom.",
      "recall_number": "Z-0012-2013",
      "product_description": "Baxter Interlink System Buretrol Solution Set with 150 mL Burette (Ball-Valve Drip Chamber); a sterile Rx IV fluid pathway device.      Product Usage:  The device is used for the administration of fluids from a container into the patient's vascular system through a vascular access device.",
      "product_quantity": "278,420 units",
      "reason_for_recall": "Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.",
      "recall_initiation_date": "20120907",
      "center_classification_date": "20121009",
      "termination_date": "20171229",
      "report_date": "20121017",
      "code_info": "product code 2C7519, all lots"
    }
  ]
}