  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hatfield",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69064",
      "recalling_firm": "Gendex Corp",
      "address_1": "2800 Crystal Dr",
      "address_2": "N/A",
      "postal_code": "19440-1944",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of Europe, Canada, Australia, China, Russia, Taiwan, Singapore, and Thailand.",
      "recall_number": "Z-0009-2015",
      "product_description": "eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly  w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm  & Tubehead, 765DC    Product Usage:  intraoral x-ray",
      "product_quantity": "651",
      "reason_for_recall": "Two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube head (full load) being installed onto the assembled articulated arm.",
      "recall_initiation_date": "20140812",
      "center_classification_date": "20141002",
      "termination_date": "20170324",
      "report_date": "20141008",
      "code_info": "Models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly  w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm  & Tubehead, 765DC; with multiple part numbers and serial numbers: see distribution list for these details."
    }
  ]
}