{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-21",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bedford",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99843",
      "recalling_firm": "Instrumentation Laboratory",
      "address_1": "180 Hartwell Rd",
      "address_2": "",
      "postal_code": "01730-2443",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.",
      "recall_number": "Z-0001-2027",
      "product_description": "HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic",
      "product_quantity": "434 units",
      "reason_for_recall": "Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.",
      "recall_initiation_date": "20260903",
      "center_classification_date": "20261009",
      "report_date": "20261021",
      "code_info": "Part Number: 0009800127.  UDI-DI: 08426951126545. Lot Number (Expiration Date): B40005 (01/16/2027), B38062 (11/01/2026).",
      "more_code_info": ""
    }
  ]
}