      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86214",
      "recalling_firm": "Covidien, LP",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and Global Distribution.",
      "recall_number": "Z-0001-2021",
      "product_description": "Endo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA  universal staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.",
      "product_quantity": "900,007 total",
      "reason_for_recall": "The device staples may not properly form upon application preventing adequate hemostasis.  Use of a product with this assembly error may result in incomplete staple formation.",
      "recall_initiation_date": "20200812",
      "center_classification_date": "20201001",
      "report_date": "20201007",
      "code_info": "Lot Numbers:   T0A108X  T7H003X  T8D008X  T8K108X  T9G060X  T0E051X  T7L023X  T8H047X   T8M079X  T7G003X  T8D007X  T8K047X  T9F132X"
    }
  ]
}