  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-02",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 39884
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92660",
      "recalling_firm": "Covidien LP",
      "address_1": "15 Hampshire St",
      "address_2": "",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - Worldwide Distribution  Foreign:  Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom",
      "recall_number": "Z-2372-2023",
      "product_description": "MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray    REF: 8888102004HP",
      "product_quantity": "685 units",
      "reason_for_recall": "Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to  full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus",
      "recall_initiation_date": "20230628",
      "center_classification_date": "20230811",
      "report_date": "20230823",
      "code_info": "GTIN: 20884521128009  LOT Numbers: 2200400154",
      "more_code_info": ""
    }
  ]
}