      "country": "United States",
      "city": "Allentown",
      "address_1": "901 Marcon Blvd",
      "reason_for_recall": "The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.",
      "address_2": "",
      "product_quantity": "63980",
      "code_info": "Batch: 61595170, 61595613, 61596536, 61598356, 61597084, 61600180, 61617168, 61616702, 61617169, 61616703, 61614599, 61616701, 61616700, 61614596, 61614598, 61619583, 61620763, 61622163, 61621828",
      "center_classification_date": "20181001",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of:  Australia, Canada, Guatemala and Kuwait.",
      "state": "PA",
      "product_description": "CE18TK ACCU-BLOC PERIFIX KIT, Material Number 332220     Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.",
      "report_date": "20181010",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "B. Braun Medical, Inc.",
      "recall_number": "Z-3362-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80811",
      "termination_date": "20200513",
      "more_code_info": "",
      "recall_initiation_date": "20180720",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}