  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Allentown",
      "address_1": "901 Marcon Blvd",
      "reason_for_recall": "The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.",
      "address_2": "",
      "product_quantity": "372",
      "code_info": "Batch: 61614720, 61622724",
      "center_classification_date": "20181001",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of:  Australia, Canada, Guatemala and Kuwait.",
      "state": "PA",
      "product_description": "CE18TIN CONT EPIDURAL SET LF, Material Number 332201     Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.",
      "report_date": "20181010",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "B. Braun Medical, Inc.",
      "recall_number": "Z-3351-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80811",
      "termination_date": "20200513",
      "more_code_info": "",
      "recall_initiation_date": "20180720",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}