  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80925",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "PO Box 610",
      "postal_code": "18034-0610",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed US nationwide and Canada.",
      "recall_number": "Z-3304-2018",
      "product_description": "Ultrasonic Bronchoscope BF-UC180F",
      "product_quantity": "4785",
      "reason_for_recall": "The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.",
      "recall_initiation_date": "20180808",
      "center_classification_date": "20180929",
      "termination_date": "20220517",
      "report_date": "20181010",
      "code_info": "all lot codes",
      "more_code_info": ""
    }
  ]
}