      "status": "Terminated",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80925",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "PO Box 610",
      "postal_code": "18034-0610",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed US nationwide and Canada.",
      "recall_number": "Z-3299-2018",
      "product_description": "Bronchofiberscope BF-P40",
      "product_quantity": "5607",
      "reason_for_recall": "The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.",
      "recall_initiation_date": "20180808",
      "center_classification_date": "20180929",
      "termination_date": "20220517",
      "report_date": "20181010",
      "code_info": "all lot codes",
      "more_code_info": ""
    }
  ]
}