{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bridgend",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80921",
      "recalling_firm": "Biomet UK Ltd.",
      "address_1": "Waterton Industrial Estate",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide in the states of AZ, MO, NJ and NM",
      "recall_number": "Z-3264-2018",
      "product_description": "HipLOC Hip Fracture Fixation System-Hiploc System 140¿ Plate 38 mm Barrel 4 Hole Plate  Item Number: 2110-140-004      Product Usage:  The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures",
      "product_quantity": "6 units",
      "reason_for_recall": "HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel",
      "recall_initiation_date": "20180809",
      "center_classification_date": "20180927",
      "termination_date": "20200512",
      "report_date": "20181003",
      "code_info": "Lot Number  3170094  3369699  3622113  3664778  3680647",
      "more_code_info": ""
    }
  ]
}