  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bridgend",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80921",
      "recalling_firm": "Biomet UK Ltd.",
      "address_1": "Waterton Industrial Estate",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide in the states of AZ, MO, NJ and NM",
      "recall_number": "Z-3262-2018",
      "product_description": "HipLOC Hip Fracture Fixation System-Hiploc System 135¿ Plate 38 mm Barrel 3 Hole Plate    Item Number: 2110-135-003    Product Usage:  The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures",
      "product_quantity": "10 units",
      "reason_for_recall": "HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel",
      "recall_initiation_date": "20180809",
      "center_classification_date": "20180927",
      "termination_date": "20200512",
      "report_date": "20181003",
      "code_info": "Lot Number:   3680822  3680823  3680824  3680825  3700273",
      "more_code_info": ""
    }
  ]
}