  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mortsel",
      "state": "",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81046",
      "recalling_firm": "Agfa N.V.",
      "address_1": "Septestraat 27",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of  CA, DC, FL, IL, IN, KS, LA, MA, MN, MO, NV, NY, OH, OR, SC and TN. Foreign distribution to Canada.",
      "recall_number": "Z-3245-2018",
      "product_description": "DX-D 600    Product Usage:  The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.",
      "product_quantity": "108",
      "reason_for_recall": "A supplier for the Agfa system noted a potential for the steel support cables of a similar device to break. This led to a mandatory review of all systems, changing the pulley and cables if any deterioration in the cables or irregular or excessive pulley wear. If the two cables that support the equipment to its roof anchor break, the equipment might fall and cause harm to the patient, or others.",
      "recall_initiation_date": "20180815",
      "center_classification_date": "20180925",
      "termination_date": "20200623",
      "report_date": "20181003",
      "code_info": "Serial Nos: A5430000015  A5430000020  A5430000030  A5430000042  A5430000045  A5430000052  A5430000053  A5430000054  A5430000057  A5430000058  A5430000060  A5430000061  A5430000063  A5430000066  A5430000067  A5430000068  A5430000074  A5430000079  A5430000081  A5430000085  A5430000089  A5430000095  A5430000096  A5430000100  A5430000105  A5430000106  A5430000107  A5430000108  A5430000119  A5430000121  A5430000124  A5430000125  A5430000130  A5430000131  A5430000132  A5430000133  A5430000137  A5430000143  A5430000144  A5430000149  A5430000156  A5430000162  A5430000165  A5430000166  A5430000172  A5430000176  A5430000177  A5430000183  A5430000184  A5430000199  A5430000206  A5430000207  A5430000215  A5430000218  A5430000222  A5430000224  A5430000225  A5430000226  A5430000228  A5430000229  A5430000234  A5430000237  A5430000238  A5430000242  A5430000246  A5430000247  A5430000248  A5430000249  A5430000263  A5430000264  A5430000272  A5430000275  A5430000276  A5430000279  A5430000281  A5430000285  A5430000293  A5430000295  A5430000296  A5430000297  A5430000298  A5430000299  A5430000300  A5430000303  A5430000304  A5430000305  A5430000306  A5430000311  A5430000312  A5430000313  A5430000320  A5430000321  A5430000322  A5430000323  A5430000324  A5430000331  A5430000332  A5430000338  A5430000350  A5430000351  A5430000354  A5430000357  A5430000358  A5430000359  A5430000360  A5430000368  A5430000369  A5430000374",
      "more_code_info": ""
    }
  ]
}