  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "6200 Jackson Rd",
      "reason_for_recall": "Screws on a Terumo¿ Advanced Perfusion  System 1 Power Manager Board (the base unit circuit board) were not torqued to specification.  It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.",
      "address_2": "",
      "product_quantity": "47 units",
      "code_info": "Serial Numbers:  0386                            1481                            1497                            1509                            1526                            1529                            1531                            1532                            1535                            1536                            1537                            1538                            1539                            1540                            1541                            1542                            1543                            1544                            1545                            1546                            1548                            1549                            1550                            1551                            1552                            1553                            1554                            1555                            1556                            1557                            1558                            1561                            1562                            1563                            1564                            1565                            1566                            1570                            1597                            1612                            1630                            1631                            1640                            1641                            1649    1547                            1560                            1559                            1567       Spare Parts:   870509   ASSEMBLY POWER MANAGER POD APS  S/N:  03130 , 03308",
      "center_classification_date": "20180925",
      "distribution_pattern": "Nationwide  Foreign:   Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam",
      "state": "MI",
      "product_description": "Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures.    Catalog Number 801763",
      "report_date": "20181003",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "recall_number": "Z-3242-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80957",
      "termination_date": "20200423",
      "more_code_info": "",
      "recall_initiation_date": "20180828",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}