{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Selzach",
      "state": "",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80863",
      "recalling_firm": "Stryker GmbH",
      "address_1": "Bohnackerweg 1",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-3240-2018",
      "product_description": "Hoffmann LRF Hexapod Software   Catalog #:  983031\t   Software Version #:   HLRF 1.0-2/002, HLRF 1.1-2/002",
      "product_quantity": "233",
      "reason_for_recall": "When creating a correction plan with the  Correct Axial First  button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.",
      "recall_initiation_date": "20180726",
      "center_classification_date": "20180922",
      "termination_date": "20200526",
      "report_date": "20181003",
      "code_info": "Version #:   HLRF 1.0-2/002, HLRF 1.1-2/002",
      "more_code_info": ""
    }
  ]
}