      "status": "Terminated",
      "city": "Saluggia, VC",
      "state": "N/A",
      "country": "Italy",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77983",
      "recalling_firm": "Sorin Group Italia SRL - CRF",
      "address_1": "Via Crescentino",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-3228-2017",
      "product_description": "Platinium VR 1210, Model Number TDF031U, UDI/GTIN 08031527015514, Implantable cardioverter defibrillator,   Biventricular or ventricular antitachycardia pacing,   Dual or single chamber arrhythmia detection    Product  Usage:  PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).",
      "product_quantity": "125 units",
      "reason_for_recall": "There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.",
      "recall_initiation_date": "20170724",
      "center_classification_date": "20170927",
      "termination_date": "20240129",
      "report_date": "20171004",
      "code_info": "All Serial Numbers"
    }
  ]
}