  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-23",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99558",
      "recalling_firm": "Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.",
      "recall_number": "Z-3205-2026",
      "product_description": "SCORPIO + MOBILE BEARING PCS\t  \t78-5-0910  \t  SYSTEM ONE BIPOLAR 40MM\t  \t6074-4022    SYSTEM ONE BIPOLAR 43MM\t  \t6074-4322  \t  SYSTEM ONE BIPOLAR 44MM\t  \t6074-4422      \t  Orthopaedic implant components manufactured from UHMWPE (polyethylene)",
      "product_quantity": "12,156 US",
      "reason_for_recall": "Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.",
      "recall_initiation_date": "20260730",
      "center_classification_date": "20260916",
      "report_date": "20260923",
      "code_info": "78-5-0910    UDI-DI 07613327366006    Lot 30582401    6074-4022    UDI-DI 07613327031775    Lots 2M5NTY   LH25R5   M36N83    6074-4322    UDI-DI 07613327031805    Lot W40E3T    6074-4422    UDI-DI 07613327031812    Lot R84K20",
      "more_code_info": ""
    }
  ]
}