  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-23",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99558",
      "recalling_firm": "Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.",
      "recall_number": "Z-3204-2026",
      "product_description": "TRIDENT 0¿ POLY INSERT 28mm ID\t  \t620-00-28D    \t  Orthopaedic implant components manufactured from UHMWPE (polyethylene)",
      "product_quantity": "12,156 US",
      "reason_for_recall": "Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.",
      "recall_initiation_date": "20260730",
      "center_classification_date": "20260916",
      "report_date": "20260923",
      "code_info": "620-00-28D    UDI-DI 07613327313581    Lots MHANP2   MHD1N0",
      "more_code_info": ""
    }
  ]
}