  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Orlando",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80761",
      "recalling_firm": "Invivo Corporation",
      "address_1": "12151 Research Pkwy Ste 200",
      "address_2": "N/A",
      "postal_code": "32826-3286",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution in the Netherlands",
      "recall_number": "Z-3192-2018",
      "product_description": "Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388",
      "product_quantity": "38",
      "reason_for_recall": "Released with an incorrect, older firmware version installed.  May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.",
      "recall_initiation_date": "20180803",
      "center_classification_date": "20180918",
      "termination_date": "20210715",
      "report_date": "20180926",
      "code_info": "System Serial Numbers:  US1814000001, US1814000002, US1814000003, US1814000004, US1814000006, US1814000007, US1814000008, US1814000011, US1814000012, US1814000013, US1814000014,   US1814000016, US1814000017, US1814000018, US1814000019, US1814000020, US1814000022, US1814000023, US1814000024, US1814000027, US1814000028, US1814000030, US1814000031,   US1814000032, US1814000033, US1814000034, US1814000037, US1814000044, US1814000046, US1814000050, US1814000053, US1814000062, US1814000066, US1814000068,and US1814000072"
    }
  ]
}