  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77417",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide in USA; Internationally to: Australia, Brazil, Canada, China,  El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.",
      "recall_number": "Z-3190-2017",
      "product_description": "Drill Bit Jacobs Chuck 2.0 mm Diameter 85 mm Length, Item Number/EDI 00480508520, Nonsterile. orthopedic surgical instrument.",
      "product_quantity": "30 units",
      "reason_for_recall": "Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly.  The anomaly may be on or below the surface and can increase the risk of instrument fracture.",
      "recall_initiation_date": "20170522",
      "center_classification_date": "20170922",
      "termination_date": "20190329",
      "report_date": "20171004",
      "code_info": "Lot Number: 63602805"
    }
  ]
}