  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77417",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide in USA; Internationally to: Australia, Brazil, Canada, China,  El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.",
      "recall_number": "Z-3182-2017",
      "product_description": "ECT Internal Fracture Fixation Quick-Connect Drill 2.0 mm Diameter 10 cm Length, Item Number/EDI 00241002000, Nonsterile. orthopedic surgical instrument.",
      "product_quantity": "418 units",
      "reason_for_recall": "Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly.  The anomaly may be on or below the surface and can increase the risk of instrument fracture.",
      "recall_initiation_date": "20170522",
      "center_classification_date": "20170922",
      "termination_date": "20190329",
      "report_date": "20171004",
      "code_info": "Lot Numbers: 63557936, 63566282, 63574978 & 63623144."
    }
  ]
}