{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Durham",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80860",
      "recalling_firm": "bioMerieux, Inc.",
      "address_1": "100 Rodolphe St",
      "address_2": "N/A",
      "postal_code": "27712-9402",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution US Nationwide and the countries of:    Turkey, Austria, Switzerland, Czech Republic, Germany, Netherlands,",
      "recall_number": "Z-3179-2018",
      "product_description": "VITEK(R) 2 AST-P640, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 418579     Product Usage:  VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth-based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.",
      "product_quantity": "2538 cartons",
      "reason_for_recall": "Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).",
      "recall_initiation_date": "20180813",
      "center_classification_date": "20180914",
      "termination_date": "20201102",
      "report_date": "20180926",
      "code_info": "Lot Numbers:  7400410103, 7400697403"
    }
  ]
}