  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80843",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states:  IN, TN.    The products were distributed to the following foreign countries:  Australia, Chile, Netherlands.",
      "recall_number": "Z-3176-2018",
      "product_description": "Gentle Thread PLGA Round Head Interference Screw, 7x20mm, Item Number 905612",
      "product_quantity": "52",
      "reason_for_recall": "The Gentle Thread PLGA Round Head, 7x20mm and the Gentle Thread PLGA Full Thread 10x30 have been commingled.",
      "recall_initiation_date": "20180720",
      "center_classification_date": "20180914",
      "termination_date": "20200604",
      "report_date": "20180926",
      "code_info": "Lot Number 611330, UDI (1)00880304008526(17)221010(10)611330"
    }
  ]
}