  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77417",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide in USA; Internationally to: Australia, Brazil, Canada, China,  El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.",
      "recall_number": "Z-3165-2017",
      "product_description": "ECT Internal Fracture Fixation Drill Bits 2.0 mm Diameter (.079 Inch) , Item Number/EDI 00231802006, Nonsterile. orthopedic surgical instrument.",
      "product_quantity": "3,640 units",
      "reason_for_recall": "Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly.  The anomaly may be on or below the surface and can increase the risk of instrument fracture.",
      "recall_initiation_date": "20170522",
      "center_classification_date": "20170922",
      "termination_date": "20190329",
      "report_date": "20171004",
      "code_info": "Lot Numbers: 63535521, 63535522, 63535523, 63535524, 63545285, 63562130, 63578154, 63578156, 63358058, 63358059, 63358060, 63358061, 63364050, 63377480, 63377481, 63377482, 63384605, 63384606, 63384607, 63398721, 63398722, 63398723, 63398724, 63398736, 63398738, 63415279, 63415280, 63415282, 63415283, 63415284, 63415286, 63426490, 63426491, 63431727, 63431728, 63431729, 63438323, 63438326, 63444503, 63444514, 63444682, 63444683, 63444684, 63444685, 63444686, 63457777, 63457946, 63457947, 63457948, 63469905, 63469907, 63510887, 63510888, 63517978, 63517997, 63518010 & 63524434."
    }
  ]
}