      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77417",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide in USA; Internationally to: Australia, Brazil, Canada, China,  El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.",
      "recall_number": "Z-3159-2017",
      "product_description": "M/DN, Intramedullary Fixation Distal Drill Pilot Tip 3.7 mm Diameter Item Number/EDI 00225503337, Nonsterile. orthopedic surgical instrument.",
      "product_quantity": "27 units",
      "reason_for_recall": "Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly.  The anomaly may be on or below the surface and can increase the risk of instrument fracture.",
      "recall_initiation_date": "20170522",
      "center_classification_date": "20170922",
      "termination_date": "20190329",
      "report_date": "20171004",
      "code_info": "Lot Number: 63505310"
    }
  ]
}