  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-23",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99637",
      "recalling_firm": "GE Medical Systems Information Technologies Inc",
      "address_1": "3114 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-3143-2026",
      "product_description": "Carescape Central Station    CSCS V3 MAS800 Desktop  Part Number 5867474-04, 5867474-13  Product code STU    CSCS V3 MAI800  Integrated   Part Number 5867474-03  Product Code STV",
      "product_quantity": "29,570 (23,632 US; 5,938 OUS)",
      "reason_for_recall": "Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.",
      "recall_initiation_date": "20260730",
      "center_classification_date": "20260911",
      "report_date": "20260923",
      "code_info": "5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU  5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU  5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV",
      "more_code_info": ""
    }
  ]
}