      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77997",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: United States (nationwide) and countries of: Angola, Australia, Botswana, Canada, Chile, Columbia, Ghana, Israel, Malaysia, Namibia, Pakistan, Panama, South Africa, Uganda, France, Switzerland, Poland, Portugal, Germany, Belgium, Finland, Czech Republic, Denmark, Netherland, Turkey, Spain, Great Britain, Romania, Brazil, Italy, Sweden, Ghana, Colombia, Lesotho and Zambia",
      "recall_number": "Z-3135-2017",
      "product_description": "AQUIOS CL Flow Cytometer System    The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.",
      "product_quantity": "213 units (34 units in the US)",
      "reason_for_recall": "Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which manifests in incorrect values being reported out of the laboratory for some patient samples.",
      "recall_initiation_date": "20170821",
      "center_classification_date": "20170919",
      "termination_date": "20230927",
      "report_date": "20170927",
      "code_info": "All software versions.; Serial numbers - BA30048,  AY20032,  BA01007,  AZ19046,  AZ12030,  BA25046,  BA05011,  AY51072,  BA01002,  AZ15044,  BA30049,  BA30050,  BA05015,  AZ19047,  BA18029,  AY47071,  AZ15038,  BA23036,  BA23035,  AZ23058,  AZ20057,  AY47063,  BA15024,  BA15023,  AZ45080,  BA01010,  AZ25070,  AZ32071,  AZ12033,  AZ25065,  AZ25068,  AY51075,  AZ19052,  BA15025",
      "more_code_info": ""
    }
  ]
}