  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mountain Home",
      "state": "AR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78085",
      "recalling_firm": "Baxter Healthcare Corp",
      "address_1": "1900 Highway 201 N",
      "address_2": "N/A",
      "postal_code": "72653-2433",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "nationwide, Canada",
      "recall_number": "Z-3133-2017",
      "product_description": "Baxter Amia Automated Peritoneal Dialysis Set with Cassette, REF 5C5479",
      "product_quantity": "164,700 units",
      "reason_for_recall": "The firm received increased customer complaints for Missing Red Line, Patient Slow Flow, Solution Slow Flow, and Inadequate Drain alerts on certain lots of the AMIA Automated Peritoneal Dialysis Set with Cassette.",
      "recall_initiation_date": "20170901",
      "center_classification_date": "20170914",
      "termination_date": "20190827",
      "report_date": "20170920",
      "code_info": "Lot numbers:  H16G28107, H16H01086, H16H04114, H16H09063, H16H13032, H16H19112, H16H26034, H16H30010, H16I02017, H16I06018, H16I08071, H16I12016, H16I14103, H16I16066, H16I19060, H16I21058, H16I23047, H16I26065, H16I30091, H16J03014, H16J05035, H16J07015, H16J09078, H16J11132, H16J14011, H16J16057, H16J18046, H16J20109, H16J22063, H16J24127, H16J26072, H16J28128, H16J30058, H16K01057, H16K03079, H16K05041, H16K07054, H16K08086, H16K10090, H16K28027, H16K30015, H16L04034, H16L11039, H16L15105"
    }
  ]
}