  "results": [
    {
      "status": "Terminated",
      "city": "Innsbruck",
      "state": "N/A",
      "country": "Austria",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77949",
      "recalling_firm": "Leonhard Lang Medizintechnik GmbH",
      "address_1": "Archenweg 56",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution. US distribution in the states of : FL & TX.  Austria, China,  France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.",
      "recall_number": "Z-3109-2017",
      "product_description": "Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion.    Product Usage:  Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only.      Used with automated external defibrillators",
      "product_quantity": "8,249 units total",
      "reason_for_recall": "Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.",
      "recall_initiation_date": "20170808",
      "center_classification_date": "20170912",
      "termination_date": "20200609",
      "report_date": "20170920",
      "code_info": "Model Number DF59N (lot numbers 41219-0778, 50403-0772, 41029-0770), DF59NC (lot numbers 50211-0974, 50402-0974, 40806-0972, 41203-0973, 50116-0975, 50527-0974). Expiration date 09/2017 to 05/31/2018."
    }
  ]
}