  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Borehamwood",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77753",
      "recalling_firm": "Stanmore Implants Worldwide Ltd.",
      "address_1": "Centennial Park",
      "address_2": "210 Centennial Avenue; CENTENNIAL PARK",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : CA, FL, GA, KS, MI, NJ, OH, PA, SC, and TX.",
      "recall_number": "Z-3104-2017",
      "product_description": "JTS External Drive Unit    JTS Distal Femoral Extendible Implant Femur is intended for the replacement of diseased or deficient bone in the distal femur.",
      "product_quantity": "15 units",
      "reason_for_recall": "Three (3) complaints were identified  in which the potential for utilization of the JTS Grower Drive in the incorrect orientation in relation to the limb entry point of the device, together with the potential to select incorrect settings for the extension procedure. As a result, the prostheses in such cases failed to extend as intended.",
      "recall_initiation_date": "20170704",
      "center_classification_date": "20170905",
      "termination_date": "20210216",
      "report_date": "20170913",
      "code_info": "Product Code MLE3; Serial No. 907-103, 907-104, 907-119, 907-123, 907-125, 907-126, 907-127, 907-028, 907-237, 907-240, 907-500, 907-501, 907-502, 907-503, 907-504."
    }
  ]
}