  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77908",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-3098-2017",
      "product_description": "Artis zee/zeego and Artis Q/Q.zen system,  x-ray, angiographic  AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis",
      "product_quantity": "986 units in USA",
      "reason_for_recall": "Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet.  Loss of coolant can results in a cooling system failure and subsequent system failure.  This leek can occur sporadically in affected systems.",
      "recall_initiation_date": "20160218",
      "center_classification_date": "20170829",
      "termination_date": "20171117",
      "report_date": "20170906",
      "code_info": "Artis zeego- Serial number: 160851  160879  160943  160484  160909  160502  160817  160999  160449  160852  160834  160972  160463  160944  160823  160994  160959  160913  160886  160961  160506  160845  160877  160488  160997  160864  160937  160965  160800  160455  160884  160486  160435  160451  160951  160431  160448  160925  160916  160814  160872  160493  160905  160470  160915  160447  160471  160481  160871  160989  160501  160869  160874  160862  160438  160860  160900  160450  160854  160878  160446  160957  160992  160856  160986  160432  160934  160947  160953  160903  160433  160917  160483  161005  160861  160889  160813  160998  160356  160914  160429  160930  160812  160460  160485  160443  160461  160955  160952  160954  160873  160912  160508  160968  160919  160940  160407  160876  160923  160885  160950  160910  160987  161000  160838  160494  160507  160487  160833  160967  160007  160825  160850  160942  160816  160482  160840  160462  160520  160505  160956  160445  160474  160428  160801  160498  160996  160509  161004  160978  160551  160896  160894  160865  160830  160807  160898  160477  160519  160469"
    }
  ]
}