      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77908",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-3097-2017",
      "product_description": "Artis zee/zeego and Artis Q/Q.zen system,  x-ray, angiographic  AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis",
      "product_quantity": "986 units in USA",
      "reason_for_recall": "Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet.  Loss of coolant can results in a cooling system failure and subsequent system failure.  This leek can occur sporadically in affected systems.",
      "recall_initiation_date": "20160218",
      "center_classification_date": "20170829",
      "termination_date": "20171117",
      "report_date": "20170906",
      "code_info": "Artis zee multi-purpose - Serial number: 157660  158170  130017  157698  158108  157657  157988  157690  158091  157616  157710  158132  157677  157615  157700  157977  157602  158183  157642  158121  157975  157969  157579  157573  157650  157671  157713  158080  130013  157928  157696  157989  157644  157648  157999  158066  158180  158032  158134  130019  157601  157992  157625  158079  157600  157604  157634  157635  158081  158104  158105  158056  157965  157570  157658  157594  158027  157706  158042  157674  157907  158043  157668  158019  157692  157936  157984  158113  157707  157708  157661  157688  158136  157987  158028  157702  158191  158138  158126  157704  158186  157667  158133  157645"
    }
  ]
}