  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77908",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-3095-2017",
      "product_description": "Artis zee/zeego and Artis Q/Q.zen system,  x-ray, angiographic  AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis",
      "product_quantity": "986 units in USA",
      "reason_for_recall": "Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet.  Loss of coolant can results in a cooling system failure and subsequent system failure.  This leek can occur sporadically in affected systems.",
      "recall_initiation_date": "20160218",
      "center_classification_date": "20170829",
      "termination_date": "20171117",
      "report_date": "20170906",
      "code_info": "Artis zee floor- Serial number: 137623  137165  137282  137618  137226  136513  136562  137118  137103  137461  137510  137193  137479  137425  137540  137532  137314  137295  137582  137592  136568  137075  137130  137188  137131  137146  136607  136579  137388  137470  124007  124015  137347  137638  137389  137595  137245  137620  137335  137617  137607  136500  136515  137369  137569  136479  137119  124051  136488  136550  136450  136408  136547  136508  136517  124039  137360  137311  137320  136586  136520  137521  137639  136548  137645  137395  137093  137018  136483  136555  137421  137551  137471  137591  137239  137438  136614  137384  137011  136567  124034  137403  137391  137296  137458  136970  137229  137112  136561  136505  124006  137244  137231  137219  137513  137224  137217  136979  137329  137328  136557  136558  137511  136503  137297  136582  136570  137567  136459  136518  136609  136594  124056  137579  137218  137518"
    }
  ]
}