  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77908",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-3092-2017",
      "product_description": "Artis zee/zeego and Artis Q/Q.zen system,  x-ray, angiographic  AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis",
      "product_quantity": "986 units in USA",
      "reason_for_recall": "Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet.  Loss of coolant can results in a cooling system failure and subsequent system failure.  This leek can occur sporadically in affected systems.",
      "recall_initiation_date": "20160218",
      "center_classification_date": "20170829",
      "termination_date": "20171117",
      "report_date": "20170906",
      "code_info": "Artis zee biplane- Serial number: 154502  154616  154536  154096  154454  154715  154789  154779  154532  131017  154157  154722  154721  154571  154074  154612  154090  154658  154686  154095  154567  154480  154093  154086  154704  153882  154727  154408  154509  154079  154462  154098  154073  154076  154792  154156  154799  154814  154783  154731  154683  153878  154565  154653  154671  154556  154406  154518  154778  154685  137650  131005  154633  154020  154058  154155  153884  154453  154708  154579  154762  154034  154018  154560  154596  154066  154663  154544  154614  154048  154800  154592  154684  154087  154085  154471  154606  154604  154456  154687  154707  154459  131000  154581  154423  154546  154052  154028  154084  154732  154709  154678  131011  154430  154501  154514  154075  154064  154562  154585  154044  154770  154035  154056  154785  154425  154082  154429  154055  154631  154476  154088  154067  154584  154610  154094  154576  154089  154041  154739  154077  154602  154097  154615  154572  154564"
    }
  ]
}