  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Orange City",
      "state": "IA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77874",
      "recalling_firm": "Med Tec Inc",
      "address_1": "PO Box 320",
      "address_2": "1401 8th St SE",
      "postal_code": "51041-7463",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of : MI, IL and TX.,  and to the countries of : Greece, Australia, Czech Republic, Hungary, Norway, Brazil, Italy, Switzerland ,UK, Austria, Slovenia, Croatia,  Poland, Germany, Japan, France, Netherland, Sri Lanka, Pakistan, Bangladesh, and Israel",
      "recall_number": "Z-3082-2017",
      "product_description": "CIVCO Arm Support, REF 106015, distributed as:    (a) Rigid Arm Support, Standard, Set Red (PN: 106015)  (b) Rigid Adj. Arm Support, Set (PN: 106020)  (c) PSR2-CR-DLR-2P Posirest-2 (PN: 106521)  (d) PB2-BLR Posiboard-2 (PN: 109030)    The Rigid Arm Supports are utilized on the Posibboard2 and Posirest assemblies, which are intended to position the patients\u0019 upper extremities during breast and lung diagnostic and radiotherapy treatment..",
      "product_quantity": "101 units",
      "reason_for_recall": "Rigid Arm Supports (PN:  106015 and 106047), which were used as part of the Posiboard-2 and Posirest, and were distributed as PN: 106015, 106047, 106521, 106522, 109030, and 109040 between Ocotober 2009 and July 2012 may have a height difference of approximately 17mm.",
      "recall_initiation_date": "20170810",
      "center_classification_date": "20170829",
      "termination_date": "20240319",
      "report_date": "20170906",
      "code_info": "Codes: M207190, M212200, M141830, M142280, M144220, M149540, M149550, M149560, M149580, M164440, M164470, M177290, M177330, M177340, M198070, M198100, M285840, M133130, M137690, M138630, M138640, M138650, M165550, M165560, M180420, M180430, M180440, M189150, M217660, M217670, M217680, M217710"
    }
  ]
}